TECHNICAL SPECIFICATIONS
ANALYSES TYPE |
Clinical Chemistry, Turbidimetric, DOA, I.S.E. |
THROUGPUT BT3500 |
up to 360 Chem. Tests/hr + up to 190 ISE Tests/hr |
THROUGHPUT BT3500S |
up to 360 Chem. Tests/hr + up to 240 ISE Tests7hr |
WATER CONSUMPTION |
1620 ml/hr ± 20% |
ISE MODULE BT3500 |
Na, K, Cl, and CO2 indirect potentiometry serum and urine |
ISE MODULE BT3500S |
Na, K, Cl, CO2 indirect potentiometry serum, diluted urine |
NUMBER OF REAGENTS |
80 Total |
REFRIGERATED REAGENT BAY |
Yes |
REACTION VOLUME |
from 180 to 400 μl |
SAMPLE VOLUME |
from 1.8 to 100 μl |
READING CUVETTES |
45 non-disposable optical glass |
SAMPLING ARMS |
2, for Serum and Reagents |
SAMPLE TRAY # OF POSITIONS |
78 Total |
TEST MODES |
Random Access, Batch, and STAT |
TEST DIRECTORY |
Up to 500 |
CUVETTE TEMP. CONTROL |
Peltier System OFF/30/32/37°C (±0.2°C) |
VACUUM PUMP SYSTEM |
External |
PHOTOMETER |
Solid State 10 Channel+Reference channel |
PHOTOMETER LINEARITY |
Uo to 3.0 O.D. |
PHOTOMETER SENSITIVITY |
± 0.001 Abs. |
DILUTER |
2: Long-Life Ceramic, 340 μl |
DILUTER ACCURACY |
± 0.1% f.s. |
DILUTER RESOLUTION |
0.1 μl |
CLOT SENSOR |
2: one for each sampling arm |
BAR CODE |
2: Reagent and Serum |
BUILT-IN QC PROGRAM |
3 Levels Known, 3 Levels Unknown |
MAINTENANCE |
Software assisted |
VETERINARY VERSION |
OPTIONAL |
COMPUTER |
Internal Windows© PC |
PROCESS ELECTRONICS |
12 microcontrollers, I² CBUS protocol |
DEVICES |
DVD, HUB USB, HD > 100 GB |
OPERATIVE SYSTEM |
Windows© XP Pro or later |
PERIPHERALS |
OPTIONAL, on request printer & UPS |
MONITOR |
Color LCD 12″ |
TOUCH-SCREEN |
Glass-protected, resistive |
KEYBOARD & House |
International |
INTERFACE |
Rs232+USB+LAN+VGA+Ethernet |
REMOTE DIAGNOSTICS (MODEM) |
Optional |
POWER SUPPLY |
Universal input: 100/240 Volt, 50-60 Hz, 590 Watt, PFC |
DIMENSIONS |
Height: 27″, Width: 40″, Depth: 23″ (68x100x58 cm) |
WEIGHT |
210 lb (95 Kg) |